Evaluation of axial length measurements obtained using three different techniques- applanation A-scan, immersion A-scan and optical biometer
DOI:
https://doi.org/10.18203/2320-6012.ijrms20230867Keywords:
Applanation, Axial length, Immersion, Optical biometryAbstract
Background: To compare measurements of axial length obtained with manual A-scan, immersion A-scan and optical biometer and assess the accuracy of different techniques of intraocular lens power calculation.
Methods: This was a prospective randomized study done in 90 patients over a period extending from November 2019 to October 2020 in the upgraded department of ophthalmology, Government Medical College Jammu. Intraocular lens power was calculated by measuring axial length by three different techniques and further the accuracy of the axial length values obtained were compared.
Results: In this study, mean axial length measured by applanation A-scan was 22.86±0.85 mm which ranged from 20.99-24.6 mm and axial length measured by immersion technique ranged from 21.5-24.9 mm with a mean AL of 22.92±0.85 mm. AL measured by non-contact method ranged from 20-25.08 mm with a mean AL of 23.1±0.93 mm. There was a statistically significant difference between all the three techniques regarding the measured AL (p=0.0004).
Conclusions: This study concludes that the optical biometry has greater accuracy than ultrasound biometry including applanation and immersion A-scan.
References
Duke-Elder S. Cataract. In: Duke-Elder S: Diseases of the Lens and Vitreous, Glaucoma and Hypotony. System of Ophthalmology. Vol-XI, 1st edn. Kimpton, London; 1969:63-64.
Yanoff M, Duker JS. Ophthalmology E-Book. Elsevier Health Sciences; 2018.
Ridley H. Intraocular acrylic lenses: a recent development in surgery of cataract. Br J Ophthalmol. 1952;36(3):113-22.
Fyodorov SN, Kolinko AI, Kolinko AI. Estimation of optical power of the intraocular lens. Vestn Oflalmol. 1967;80:27-31.
Fyodorov SN, Galin MA, Linksz A. A calculation of the optical power of intraocular lenses. Invest Ophthalmol. 1975;14:625-8.
Hillman JS. 1982: Ophthalmic ultrasound biometry and the calculation of intraocular lens power. Ultrasound Med Biol. 1982;8(Suppl 1):78.
Ademola-Popoola DS, Nzeh DA, Saka SE, Olokoba LB, Obajolowo TS. Comparison of ocular biometry measurements by applanation and immersion A-scan techniques. J Curr Ophthalmol. 2015;27:110-4.
Shammas HJ. A comparison of immersion and contact techniques for axial length measurement. J Am Intraocul Implant Soc. 1984;10(4):444-7.
Nakhli FR. Comparison of optical biometry and applanation ultrasound measurements of the axial length of the eye. Saudi J Ophthalmol. 2014;28(4):287-91.
Gaballa SH, Allam RS, Abouhussein NB, Raafat KA. IOL master and A-scan biometry in axial length and intraocular lens power measurements. Delta J Ophthalmol. 2017;18:13-9.
Kongsap P. Comparison of a new optical biometer and a standard biometer in cataract patients. Eye Vis. 2016;3:27.
Solanki H, Patel D, Chabbra A. Comparitive study of pre-operative IOL power calculation by IOL master, immersion and non-immersion techniques (A-scan, Manual keratometer). Int J Sci Res. 2013;4(9).