Study of automated hematology analyzer’s scatterplot patterns in white blood cell disorders
DOI:
https://doi.org/10.18203/2320-6012.ijrms20240524Keywords:
Hematology analyzer, Leukemia, Scatterplots, White blood cell disordersAbstract
Background: Automated hematology analyzers produce scattergrams that can be used as screening tool for various hematological conditions and efficiently shorten turnaround times. Aim was to study scattergram patterns of various while blood cell disorders and assess their efficacy compared to a peripheral blood smear for diagnosis of various disorders.
Methods: Scattergram findings generated by UniCel® DxH 800 automated hematology analyzer, a 5-part differential analyzer. The graphic displays have been compiled over a period of 3 months from blood samples received for CBC. Samples that the counter flagged as abnormal for white blood cell were chosen. Based on the scatterplots, a preliminary diagnosis was formed. It was compared with the peripheral blood smear (PBS) findings which were taken as the gold standard.
Results: The scatterplots showed unique patterns for various disorders on the basis of location, shape, size, density of the cells and their clustering. The scattergram analyser showed 90% sensitivity and 88% specificity for diagnosing hematological disorders. A 97-100% accuracy rate was reported showing excellent correlation between PBS result and WBC parameter result in cell counter analyzers.
Conclusions: Not all cases of haematological malignancy exhibit cytopenias or cytosis at initial presentation. Therefore, these scatter plots offer helpful information that prompts a hematopathologist to suspiciously screen the peripheral smear in cases with normal counts. Scattergram analysis suspects a diagnosis earlier than peripheral smear examination. Given their strong correlation with a variety of WBC disorders and confirmed by PBS, WBC scatterplots can be used as a screening tool.
References
Beckman coulter unicel® dxh 800 coulter® cellular analysis system Manual. Available from: https://www.beckmancoulter.com/products/hematology/dxh-800. Accessed on 20 August 2023.
Gupta M, Chauhan K, Singhvi T, Kumari M, Grover RK. Useful information provided by graphic displays of automated cell counter in hematological malignancies. J Clin Lab Anal. 2018;32:e22392.
Asad S, Ahmed I, Ali N. Utility of peripheral film findings and its correlation with automated analyzer- an audit from tertiary care hospital. J Lab Phys. 2017;9:1-4.
Ningombam A, Acharya S, Sarkar A, Kumar K, Chopra A. Scattergram patterns of hematological malignancies on sysmex XN-series analyzer. J Appl Hematol. 2021;12(2):83-9.
Adewoyin AS. Peripheral blood film-a review. Ann Ibadan Postgrad Med. 2014;12(2):71-9.
Lokwani DP. The ABC of CBC: Interpretation of complete blood count and histograms. 1st edn. JP Medical Ltd; 2013:127-130.
Kawthalkar SM. Essentials of clinical pathology. 1st edn. Jaypee Brothers Medical Ltd.; 2010:323-326
Gupta T, Basu D. Utility of scatterplot patterns of automated hematology analysers in white blood cell disorders- a comparative study with peripheral blood smears. Asian J Med Sci. 2020;11(5):29-34.
Mishra S, Kumar K, Panigrahi A, Das P, Padhi S, Chhabra G. The utility of leucocyte cell population data and scattergram in rapid identification of acute promyelocytic leukemia. Blood. 2020;136:19-20.
Juthani R, Basu D. Use of scatterplot patterns derived from automated haematological analysers in the diagnosis of acute febrile illnesses. J Clin Diagn Res. 2020;14(2).
Hoyer JD, Fisher CP, Soppa VM, Lantis KL, Hanson CA. Detection and classification of acute leukemia by the Coulter STKS Hematology Analyzer. Am J Clin Pathol. 1996;106(3):352-8.
Choccalingam C. Volume, conductance, and scatter parameters of neoplastic and nonneoplastic lymphocytes using Coulter LH780. J Lab Phys. 2018;10:85-8.