DOI: http://dx.doi.org/10.18203/2320-6012.ijrms20194990

A comparison of three different gauge Quincke’s spinal needles in patients undergoing elective surgery under spinal anaesthesia: a prospective randomized study

Kiranpreet Kaur, Suresh K. Singhal, Anu Mehla

Abstract


Background: The present study was conducted to compare three different Quincke’s spinal needles i.e. 23, 25 and 26 gauge in patients undergoing elective surgery under spinal anaesthesia (SA).

Methods: The prospective randomized double-blind study was conducted on 150 male patients in the age group of 18-50 yrs., having physical status class I to II, scheduled for elective surgery under SA. Patients were randomly divided into three groups comprising 50 patients each. SA was administered using Quincke’s spinal needles of 23, 25 and 26 gauge in group 1, 2 and 3 respectively. Ease of insertion, number of attempts and time of appearance of CSF and incidence of PDPH was recorded in all the patients.

Results: Ease of insertion was graded easy in 98%, 84% and 82% in group 1, 2 and 3 respectively. First attempt success rate was highest in group 1, (98%). Meantime for appearance of CSF beyond hub was maximum in group 3 i.e. 14.60±2.56 sec. Mild PDPH was reported in 6% and 2% patients after 24 hrs in group 1 and group 2 respectively.

Conclusions: Finer spinal needle proved to be more dependable in generating less traumatic effect on the dura and preventing PDPH but are technically more difficult thus decreasing first attempt success rate.


Keywords


Needle gauge, Post-dural puncture headache, Quincke’s needle, Spinal needles, Spinal needle technical difficulty

Full Text:

PDF

References


Tsen LC, Hepner DL. Needles used for spinal anesthesia. Expert Rev. Med. Devic. 2006;3(4):1-10.

Kang SB, Goodenough DE, Lee YK, Olson RA, Borshoff JA, Furlano RMM, et al. Comparison of 26 G and 27 G needles for spinal anaesthesia for ambulatory surgery patients. Anesthesiol. 1992;76:734-8.

Mukundan M, Ninan R, Ninan M. A comparative study between 25 gauge quincke’s and Whitacre spinal needle in spinal anaesthesia in obstetrics patients. Intern J scient res. 2018;7:615-7.

Tarkilla P, Huhtala J, Salminen U. Difficulties in spinal needle use, Insertion characteristics and failure rates associated with 25, 27 and 29 gauge Quincke type spinal needles. Anaesth. 1994;49:723-5.

Oberhofer D, Jokic A, Skok I, Skuric J, Vukelic M, Sklebar I, et al. Incidence and clinical significance of postdural puncture headache in young orthopaedic patients and parturients. Periodi Biolo. 2013;115:203-8.

Shah VR, Bhosale GP. Spinal anaesthesia in young patients: evaluation of needle gauge and design on technical problems and postdural puncture headache. South Afr J Anaesthesiol Analg. 2010;16(3):324-8.

Carson DF, Serpell MG. Clinical characteristics of commonly used spinal needles. Anaesth. 1995;50(6):523-5.

Sitzman BT, Uncles DR. The effects of needle type, gauge, and tip bend on spinal needle deflection. Anesth Analg. 1996;82(2):297-301.

Bajwa BS, Cheema HK, Kaur K. Comparison of size of two spinal needles in post dural puncture headache: a retrospective study. J Evol Medi Dental Sci. 2015;4(100):499-501.

Mohammed EL, Shal SME. Efficacy of different size Quincke spinal needles in reduction of incidence of postdural puncture headache in caesarean section. Egypt J Anaesth. 2016;33:53-8.

Tang L, Tang L, Li S, Huang S, Chen L, Zhang L. Spinal anaesthesia for emergency caesarean section better using 25 gauge pencil point needle or 22 gauge cutting needle: a single centre prospective study. Int J Clin Exp Med. 2017;10:12293-300.

Syed S, Qayoom N, Naaz S, Mushtaq K, Mir A H, Bijli AH et al. Comparison of postdural puncture headache - incidence and severity in obstetric patients after spinal anaesthesia for caesarean section with 25 G and 27 G Quincke needle. Int J Res Med Sci. 2017;5:596-600.

Kumar V, George B, Dixit N. Incidence of postdural puncture headache following spinal anaesthesia in caesarean section: a comparison of 25G and Quincke and 25G Whitacre needle. Glo J Res Analy. 2018;7:5-7.

Das Neves JFNP, Monteiro GA, Almeida JRD, Brun A, Anna RSS, Duarte ES. Spinal anaesthesia with 27 G and 29 G Quincke and 27 G Whitacre needles- technical difficulties, failures and headache.Rev Brasil d Anestesiol. 2001;51:3:196-201.

Cook TM, Counsell D, Wildsmith JA. Major complications of central neuraxial block: report on the Third National Audit Project of the Royal College of Anaesthetists.Br J Anaesth. 2009;102:179-90.

Joshi V. Comparitive Study of 22, 23 and 25 Gauge Quincke Needles for Post Dural Puncture Headache (PDPH). Ind J Res. 2016;5:47-9.

Babu DD, Chandar DD, Prakash CS, Balasubramanian S, Kumar KS. Evaluation of Post Dural Puncture Headache Using Various Sizes of Spinal Needles. Int J Sci Stud. 2015;3:9-13.

Ayub F, Ahmad A, Aslam KZ, Saleem I. Frequency of headache with 25G or 27G quincke needles after spinal anesthesia in patients undergoing elective cesarean section. Anaesth, Pa & InteN Ca. 2017;21:170-3.

Irkal JN, Reddy SV, Krishan D, Bhardwaj A.Incidence of post-dural puncture headache: A comparison of quinckes’ versus whitacres’ spinal needles. Karn Anaesth J. 2016;2:81-5.